Performance tests using the Lexsyg luminescence reader |
| |
Authors: | Johanna Lomax Sebastian Kreutzer Markus Fuchs |
| |
Institution: | 1. Department of Geography, Justus-Liebig-University Giessen, Senckenbergstr. 1, 35390, Giessen, Germany 2. IRAMAT-CRP2A, Université Bordeaux Montaigne, Maison de l’Archéologie, Esplanade des Antilles, 33607, Pessac Cedex, France
|
| |
Abstract: | This study presents test measurements on a Lexsyg luminescence reader, acquired by the Giessen luminescence group. The reader is of the type Standard, hence designed for routine determination of palaeodoses using quartz or K-feldspars. The tests include measurements of the stimulation powers, preheat temperatures, OSL- and TL-curves, as well as dose recovery tests and calibration measurements using highly sensitised calibration quartz. A comparison of D e values determined using a Lexsyg reader with single grain D e-values, which have been previously obtained on a Risø reader, is also presented. The results imply that the Lexsyg reader is a highly reliable measurement device with high reproducibility, yielding instrumental uncertainties of around 0.3%. |
| |